TD-4257
Blood Glucose Meter
- Individual blood glucose target range
- 5-color indicator
Brown: <54 mg/dL
Green: Within target range
Red: Out of target range
Yellow: Out of target range
Orange: >250 mg/dL
Meter | Ketone Warning | Yes |
Communication | Strip port | |
Power Source | 2 x AAA | |
Memory Capacity | 500 sets | |
Day Average | 7, 14, 21, 28, 60, 90 days for blood glucose | |
Daily Alarm | 4 daily alarms | |
Dimension | 97 (L) x 53.7 (W) x 21.7 (H)mm | |
Weight | 81g (without battery) | |
Operating Condition | 10ºC (50ºF) ~ +40ºC (104ºF), 10% ~ 85% R.H. | |
Storage Condition | -20 ºC ~ +60 ºC (Meter); 2ºC ~ 30ºC (Strip) | |
Glucose | Enzyme Type | GDH-FAD |
Sample Size | 0.5 μL | |
Reaction Time | 5 seconds | |
Measurement Range | 20 ~ 650 mg/dL (1.1~36.1mmol/L) | |
Hematocrit Range | 20% ~ 65% | |
Precision | SD < 5 mg/dL (0.278 mmol/L) at < 100 mg/dL (5.55 mmol/L); CV < 5% at 100 mg/dL (5.55 mmol/L) |
|
Accuracy | ±15 mg/dL if < 100 mg/dL; ±15% if ≥ 100 mg/dL |
|
Package | Vial pack / Single foil pack | |
Ketone | Enzyme Type | HBD |
Sample Size | 1.0 µL | |
Reaction Time | 10 seconds | |
Measurement Range | 0.1 ~ 8.0mmol/L | |
Hematocrit Range | 20% ~ 65% | |
Precision | ≤ 1mmol/L, SD < 0.1mM; > 1mmol/L, CV < 7.5% |
|
Accuracy | ≤ 1 mmol/L, SD < 0.1 mM; > 1 mmol/L, CV < 7.5% |
|
Package | Vial pack / Single foil pack |