TD-4141
Multi-Parameter Meter
- Before / After meal (AC/PC) presettable
- High / Low indicator for blood glucose
- Strip expiry date reminder
- Strip light
Part No. | TD-4141A |
Communication | Strip Port Communication |
Power Source | AAA x 1 |
Memory Capacity | 1000 sets |
Dimension | 97.8 (L) x 33.5 (W) x 18 (H) mm |
Weight | 30g (without battery) |
Operating Condition | 10°C (50°F) ~ +40°C (104°F) |
Storage Condition | Meter: -20ºC (-4ºF) ~ +60ºC (140ºF) Strip: 2ºC (35.6ºF) ~ 30ºC (86.0ºF) |
Glucose | |
Enzyme Type | GDH-FAD |
Sample Size | 0.5 µL |
Reaction Time | 5 seconds |
Measurement Range | 10 ~ 800 mg/dL (0.56 ~ 44.4 mmol/L) |
Hematocrit Range | 0% - 70% |
Precision | ±15 mg/dL if < 100 mg/dl; ±15% if ≥ 100 mg/dL SD < 5 mg/dL (0.278 mmol/L) at < 100 mg/dL (5.55 mmol/L); |
Accuracy | CV < 5% at 100 mg/dL (5.55 mmol/L) |
Package | Vial pack / Single foil pack |
Uric Acid | |
Sample size | 0.5 µL |
Reaction time | 15 seconds |
Measurement Range | 3 ~ 20 mg/dL (178 ~ 1190 µM) |
Hematocrit Range | 20% - 60% |
Precision | ≦5 mg/dL, SD<0.5 mg/dL; 5 mg/dL, CV<7.5% (≦297 µM, SD<30 µM; >297 µM, CV<7.5%) (≦0.29 mM, SD<0.03 mM; >0.29 mM, CV<7.5%) |
Package | Single foil pack |
Total Cholesterol | |
Sample size | 3.0 µL |
Reaction time | 60 seconds |
Measurement Range | 100 ~ 400 mg/dL |
Hematocrit Range | 20% - 60% |
Precision | CV < 7.5% |
Package | Single foil pack |
Accessories | |
ACC-001 | Bluetooth dongle |