4G Cellular BGM
- LTE 4G Connectivity, Real-time cloud sync for effortless data access.
- Ultra-Low Power, Long-lasting performance for continuous monitoring.
- Remote Monitoring, Easy doctor tracking for better health management.
For TD-4141 Wireless Blood Glucose Monitor, it’s an IoT-based system designed on blood glucose meter for presenting real-time glucose monitoring and tracking.
TD-4141 Wireless Blood Glucose Monitor supports Narrow Band Internet of Thing (NB-IoT) connection.
| Enzyme Type | GDH-FAD |
| Sample Size | 0.5 µL |
| Reaction Time | 5 seconds |
| Measurement Range | 20 ~ 650 mg/dL (1.1 ~ 36.1 mmol/L) |
| Hematocrit Range | 20% ~ 65% |
| Precision | SD < 5 mg/dL (0.278 mmol/L) at < 100 mg/dL (5.55 mmol/L); CV < 5% at 100 mg/dL (5.55 mmol/L) |
| Accuracy | ±15 mg/dL if < 100 mg/dL; ±15% if ≥ 100 mg/dL |
| Communication | 4G |
| Ketone Warning | Yes |
| Power Source | Rechargeable Li-ion Battery |
| Memory Capacity | 500 sets |
| Day Average | 7, 14, 21, 28, 60, 90 days for blood glucose |
| Daily Alarm | 4 daily alarms |
| Dimension | 102 (L) x 64 (W) x 29.5 (H) mm |
| Weight | 81g (without battery) |
| Operating Condition | 10°C (50°F) ~ 40°C (104°F), 10% ~ 85% R.H. |
| Storage Condition | Meter: -20°C (-4°F) ~ 60°C (140°F); Strip: 2°C (35.6°F) ~ 30°C (86°F) |
| Package | Vial pack / Single foil pack |
| 4G MODULE | |
| Part. No. | EXS62-W |
| Description | LTE 4G Module for Global Deployment |
| Frequency Bands | LTE Cat NB1/2: 600 (Bd71), 700 (Bd12, Bd13, Bd28, Bd85), 800 (Bd18, Bd19, Bd20, Bd26), 850 (Bd5), 900 (Bd8), AWS-3 (Bd66), AWS-1 (Bd4), 1800 (Bd3), 1900 (Bd2, Bd25), 2100 (Bd1) LTE Cat M1: 700 (Bd12, Bd13, Bd28, Bd85), 800 (Bd18, Bd19, Bd20, Bd26, Bd27), 850 (Bd5), 900 (Bd8), AWS-3 (Bd66), AWS-1 (Bd4), 1800 (Bd3), 1900 (Bd2, Bd25), 2100 (Bd1) |
| SHIPPING INFO | |
| Units Per Carton | 40 |
| Carton Size | 49.8 (L) x 35.8 (W) x 36 (H) cm |
| Cross Weight Per Carton | 12.8 kg |