TD-4239
Blood Glucose Meter
- Ketone warning
- RS232 / Strip port communication
- Neonatal measurement available
- Before / After meal (AC/PC) presettable
Enzyme Type | GDH-FAD |
Sample Size | 1.0 µL |
Reaction Time | 5 seconds |
Measurement Range | 10~600mg/dL (0.5~33.3mmol/L) |
Hematocrit Range | 0% - 70% |
Precision | SD < 5mg/dL (0.278mmol/L) at < 100mg/dL (5.55mmol/L); CV < 5% at 100mg/dL (5.55mmol/L) |
Accuracy | ±15mg/dL if < 100mg/dL; ±15% if ≧ 100mg/dL |
Ketone Warning | Yes |
Communication | RS232 |
Power Source | CR2032 x 1 |
Memory Capacity | 400 sets |
Day Average | 7,14,21,28,60,90 Days |
Daily Alarm | 4 daily alarms |
Dimension | 95 (L) x 49.0 (W) x 14.0 (H) mm |
Weight | 42 g (without battery) |
Operating Condition | 10ºC(50ºF) ~ 40ºC(104ºF), below 85% R.H. |
Storage Condition | -20ºC (-4ºF) ~ +60ºC (140ºF) (Meter) 2ºC(35.6ºF)~32ºC(89.6ºF) (Strip) |
Units Per Carton | 30 |
Carton Size | 476 (L) x 370 (W) x 400 (H) mm |
Cross Weight Per Carton | 10.5 kg |