Diabetic Meter
TD-4279
- Bluetooth / Strip port communication
- Before / After meal (AC/PC) presettable
- Strip expiry date reminder
- 1000 memory
Diabetic Meter TD-4279 is intended to be used for the quantitatively measure the blood glucose and β-ketone in fresh capillary whole blood.
Diabetic Meter TD-4279 (Diabetic Monitor) have automatic test strip recognition function.
Using Diabetic Meter to detect your ketone level when during ketone diet.
Meter | Ketone Warning | Yes |
Communication | Strip port (TD-4279C) / Bluetooth (TD-4279D) | |
Power Source | 2 x AAA | |
Memory Capacity | 1000 sets | |
Day Average | 7, 14, 21, 28, 60, 90 days for blood glucose | |
Daily Alarm | 4 daily alarms | |
Dimension | 97 (L) x 62.2 (W) x 28.9 (H) mm | |
Weight | 67.5 g (without battery) | |
Operating Condition | 10°C ~ 40°C, below 85% R.H. | |
Storage Condition | -20 ºC ~ +60 ºC (Meter); 2ºC ~ 32ºC (Strip) | |
Glucose | Enzyme Type | GDH-FAD |
Sample Size | 1.0 µL | |
Reaction Time | 5 seconds | |
Measurement Range | 10 - 700 mg/dL (0.56 - 38.89 mmol/L) | |
Hematocrit Range | 0% - 70% | |
Precision | CV < 5% if > 100 mg/dL; SD < 5 mg/dL if ≤ 100 mg/dL | |
Accuracy | ±15 mg/dL if ≤ 100 mg/dL; ±15% if > 100 mg/dL | |
Package | Vial pack | |
Ketone | Enzyme Type | HBD |
Sample Size | 1.0 µL | |
Reaction Time | 10 seconds | |
Measurement Range | 0.1 ~ 8.0mmol/L | |
Hematocrit Range | 10% - 70% | |
Precision | ≤ 1 mmol/L, SD < 0.1 mM; > 1 mmol/L, CV < 7.5% | |
Accuracy | ≤ 1 mmol/L, SD < 0.1 mM; > 1 mmol/L, CV < 7.5% |
|
Package | Single foil pack | |
Accessories | ACC-01 | Bluetooth dongle |
ACC-02 | USB cable |