Diabetic Meter TD-4279 is designed to help diabetes patients keep track of their blood sugar levels well.

LCD Backlight Icon
LCD Backlight
QC Identification
QC Identification
Strip Ejection Icon
Strip Ejection
Display HCT result

Diabetic Meter

TD-4279

  • Bluetooth / Strip port communication
  • Before / After meal (AC/PC) presettable
  • Strip expiry date reminder
  • 1000 memory

Diabetic Meter TD-4279 is intended to be used for the quantitatively measure the blood glucose and β-ketone in fresh capillary whole blood.

Diabetic Meter TD-4279 (Diabetic Monitor) have automatic test strip recognition function.

Using Diabetic Meter to detect your ketone level when during ketone diet.

Specifications

MeterKetone WarningYes
CommunicationStrip port (TD-4279C) / Bluetooth (TD-4279D)
Power Source2 x AAA
Memory Capacity1000 sets
Day Average7, 14, 21, 28, 60, 90 days for blood glucose
Daily Alarm4 daily alarms
Dimension97 (L) x 62.2 (W) x 28.9 (H) mm
Weight67.5 g (without battery)
Operating Condition10°C ~ 40°C, below 85% R.H.
Storage Condition-20 ºC ~ +60 ºC (Meter); 2ºC ~ 32ºC (Strip)
Glucose Enzyme TypeGDH-FAD
Sample Size1.0 µL
Reaction Time5 seconds
Measurement Range10 - 700 mg/dL (0.56 - 38.89 mmol/L)
Hematocrit Range0% - 70%
PrecisionCV < 5% if > 100 mg/dL; SD < 5 mg/dL if ≤ 100 mg/dL
Accuracy±15 mg/dL if ≤ 100 mg/dL; ±15% if > 100 mg/dL
PackageVial pack
KetoneEnzyme TypeHBD
Sample Size1.0 µL
Reaction Time10 seconds
Measurement Range0.1 ~ 8.0mmol/L
Hematocrit Range10% - 70%
Precision≤ 1 mmol/L, SD < 0.1 mM; > 1 mmol/L, CV < 7.5%
Accuracy≤ 1 mmol/L, SD < 0.1 mM;
> 1 mmol/L, CV < 7.5%
PackageSingle foil pack
AccessoriesACC-01Bluetooth dongle
ACC-02USB cable